Detailed Product Description
Measuring Principle: Scattering Light Method Measuring Range/Resolution: 0~1000NTU/0.1NTU /0.1℃ 0~100NTU /0.1NTU /0.1℃ 0~20NTU /0.01NTU/0.1℃
Accuracy: ±5% Or ±3NTU(0~1000NTU) ±3% Or ±2NTU(0~100NTU) ±3% Or±1.5NTU(0~20NTU) ±0.5℃ Calibration Mode: Two-point Calibration
Temperature Compensation: Automatic Temperature Compensation(Pt1000) Output Mode: RS-485( Modbus /RTU)
High Light:

Two point calibration online turbidity sensor

,

Scattering light method online turbidity sensor

,

Pt1000 water turbidity sensor

KNT601 Online Turbidity Sensor
 
1.Working Principle
KNT601 integrated online turbidity sensor is designed and manufactured using the principle of scattered light turbidity measurement. When a beam of light is incident on a water sample, the light is scattered by the turbidity substance in the water sample, and the intensity of the scattered light in the direction perpendicular to the incident light is measured and compared with the internal calibration value to calculate the turbidity in the water sample. Degree, linearized to output the final value.
2.Feature

  • 90° angle scattered light principle, built-in temperature sensor
  • Support RS-485, Modbus/RTU protocol
  • Optical fiber structure, strong resistance to external light interference
  • Infrared LED light source with high stability
  • IP68 protection, water depth within 20 meters
  • Convenient, fast, stable and easy to maintain

3.Technical Specifications

Model KNT601
Measuring principle Scattering light method
Measuring range/Resolution0~1000NTU/0.1NTU /0.1℃
0~100NTU /0.1NTU /0.1℃
0~20NTU /0.01NTU/0.1℃
Accuracy

±5% or ±3NTU(0~1000NTU)
±3% or ±2NTU(0~100NTU)
±3% or±1.5NTU(0~20NTU)
±0.5℃

Calibration modeTwo-point calibration
temperature compensation Automatic temperature compensation(Pt1000)
Output modeRS-485( Modbus /RTU)
Working conditions0~50℃,<0.2MPa
Storage temperature-5~65℃
Installation modeImmersive installation, 3/4NPT thread
Cable length5 meters, other lengths customizable
power dissipation<0.3 W@12 V
source12~24 VDC ±10%
levels of protection IP68
 

 

4.Dimensional Drawing

Pt1000 Online Turbidity Sensor Scattering Light Method 0
 
 
Note:The sensor connector is m16-5 core waterproof connector male.

5.Installation and Electrical Connection

5.1 Install

Installation distance requirements: keep above 5 cm with side wall and above 10 cm with bottom.

5.2 Electrical Connection

  • Red wire-power cord(12~24 V)
  • Black wire-ground wire(GND)
  • Blue Line-485 A
  • White Line-485 B
  • Bare wire-shield line

After the wiring is completed, it should be carefully checked to avoid the wrong connection before the power is turned on.

Cable specification: considering that the cable is immersed in water (including sea water) or exposed to air for a long time, the cable has a certain corrosion resistance. All interfaces of cable outer diameter Φ6mm, have been waterproof.

6.Maintenance

6.1 Maintenance Procedures and Methods

6.1.1 Maintenance Schedule

The cleanliness of the measuring window is very important for maintaining accurate readings.

Maintenance taskRecommended maintenance frequency

Calibrate sensors

(If required by the competent authority)

According to the maintenance schedule required by the competent department
 

6.1.2 Maintenance Method

  • Sensor outer surface: clean the outer surface of the sensor with tap water, if there is still debris residue, wipe with wet soft cloth, for some stubborn dirt, you can add some household washing liquid to tap water to clean.
  • Check the cable of the sensor: the cable should not be tightened when it is working properly, otherwise it is easy to break the wire inside the cable and make the sensor unable to work properly.
  • Check the sensor measurement window if there is any dirt, cleaning brush is normal.

6.1.3 Matters Need Attention
The probe contains sensitive optical and electronic components. Make sure the probe is not subjected to severe mechanical impact. There are no components inside the probe that need to be maintained by the user.

6.2 Calibration of Sensors

a) Zero calibration: take proper amount of zero turbidity solution with large beaker, put the sensor vertically in the solution, the front end of the sensor is at least 10 cm from the bottom of the beaker, and the zero calibration will be carried out after the value is stabilized for 3 minutes. The instructions refer to the appendix.

b) Slope calibration: the sensor probe is placed in the standard solution, the front end of the sensor is at least 10 cm from the bottom of the beaker, and the slope calibration is carried out after 3 ≤ 5 minutes of numerical stability. The instructions refer to the appendix.

6.3 Frequently Asked Questions

WrongProbable CauseSolution
The operating interface cannot connect or does not display the measurement resultsThe measured value is too high, too low, or the numerical value remains unstable.Reconnect the controller and cable.
Cable failurePlease contact us.
The measured value is too high, too low, or the numerical value remains unstable.The sensor window is attached to the external object.Clean the window surface of the sensor.
 

7.Quality and Service

7.1 Quality Assurance

  • The quality inspection department has a standard inspection procedure, with advanced and complete detection equipment and means, and according to the procedure inspection, the product is subjected to 72-hour aging experiment and stability experiment, so that a non-conforming product is not allowed to leave the factory.
  • The consignee shall refund directly the product batches with a failure rate of 2%, and all expenses incurred shall be borne by the supplier. Consider the standard reference to the product description provided by the supplier.
  • Ensure the quantity of goods and the speed of shipment.

7.2 Spare Parts

This product includes:

  • One transmitter
  • One copy of explanation
  • One certificate of quality

 

7.3 After-sales Service Commitment

The Company has provided the local after-sales service within one year from the date of sales, but does not include the damage caused by improper use. If it is necessary to repair or adjust it, please return it, but the freight is required to be self-contained. When it is returned, it shall be confirmed that the package is good to avoid damage during transportation. The Company will repair the damage of the instrument free of charge.
 
 
 

Contact Us

Website www.kacise.com
Phone +86-17719566736
Email sales@kacise.com
Location Tangyan South Road, High-tech Zone, Xi'an City, Shaanxi Province, China
Pt1000 Online Turbidity Sensor Scattering Light Method 1
 

Contact Details
Xi'an Kacise Optronics Co.,Ltd.

Contact Person: Ms. Evelyn Wang

Tel: +86 17719566736

Fax: 86--17719566736

Send your inquiry directly to us (0 / 3000)

Contact

Address: i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.

Factory Address:i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.