Many organic substances dissolved in water absorb ultraviolet light. Therefore, by measuring the absorption of these organic substances by ultraviolet light at a wavelength of 254 nm, the total amount of dissolved organic pollutants in water can be accurately measured. The KCD602 online COD sensor uses two light sources, one 254 nm ultraviolet light and one 365 nm ultraviolet reference light, which can automatically compensate the optical path attenuation and eliminate the interference of the granular suspended matter enamel to a certain extent, thus achieving more stable and reliable.
| Model | KCD602 |
| Measuring principle | Dual wavelength ultraviolet absorption method |
| COD Measuring range | 0-100 mg/L equiv. KHP |
| COD Precision | ±5%F.S. |
| COD Resolution | 0.1 mg/L |
| Turbidity range | 0-100 NTU |
| Turbidity accuracy | ±5%F.S. |
| Turbidity resolution | 0.1 NTU |
| Calibration method | Two-point calibration |
| Output method | RS-485 bus, Modbus/RTU protocol |
| Power supply | 12 VDC±10% |
| Power consumption | <0.5 W |
| Working conditions | 0~45℃,<0.2 MPa |
| Storage temperature | -5~65℃ |
| Protection level | IP68 |
| Installation method | Immersion installation |
| Cable length | 5 meters, other lengths can be customized |
| Sensor housing material | 316L stainless steel |
![]()
When the sensor is attached, the sensor caused by water flow should be prevented from hitting the wall or other water conservancy facilities. If the water flow is very urgent, fix the sensor.
![]()
Install the sensor to a depth of no more than 2 meters from the water surface.
![]()
Taking into account the fluctuation of the water level, the sensor is immersed in the lowest possible water level below 30 cm.
![]()
The sensor is placed in a position where there are no bubbles in the water.
![]()
After wiring is completed, it should be carefully checked to avoid damage to the sensor caused by incorrect wiring.
Cable specification:Considering that the cable is immersed in water (including sea water) for a long time or exposed to the air, the cable has a certain degree of corrosion resistance. The outer diameter of the cable is Φ6 mm and all interfaces are waterproof.
6.1.1 Maintenance Schedule
To ensure accurate measurements, cleaning is important, and regular sensor cleaning helps data stability.
| Maintenance task | Recommended maintenance frequency |
| Calibrate the sensor (if required by the competent authority) | According to the maintenance schedule required by the competent authority |
| Maintain and check the self-cleaning brush | Return to the factory every 18 months for inspection and maintenance of self-cleaning brush |
Note: The maintenance frequency in the above table is only a recommendation. Please maintain the sensor according to the actual usage of the sensor.
6.1.2 Maintenance Method
1)Sensor outer surface: Clean the outer surface of the sensor with tap water. If there is still debris left, wipe it with a soft, damp cloth. For some stubborn dirt, add some household washing liquid to the tap water to clean it.
2)Check the cable of the sensor: the cable should not be tightened during normal operation. Otherwise, the cable inside the cable may be broken, causing the sensor to malfunction.
3) Check if the sensor's measurement window is dirty and the cleaning brush is normal.
4) Check the sensor's cleaning brush for damage.
5) For 18 months of continuous use, it is necessary to return to the factory to replace the dynamic sealing device.
1.3 Precautions
The probe contains sensitive optical and electronic components. Make sure the probe is not subject to severe mechanical shock. There are no parts inside the probe that require user maintenance.
6.2.1 Turbidity Calibration:
6.2.2 COD Calibration:
6.2.3 Calibration (2-point calibration)
Place the sensor in a 5mg/L COD solution and verify that all light paths are submerged under water >2cm with no air bubbles. Perform a zero calibration according to the appendix command.
Place the sensor in a 20mg/L COD solution and verify that all light paths are submerged under water >2cm with no air bubbles. Perform a zero calibration according to the appendix command.
| Problem | Possible reason | Solution |
|
The operation interface cannot be connected Or do not display measurement results |
Controller and cable connection error | Reconnect controller and cable |
| Cable failure | Please contact us | |
| The measured value is too high, too low or the value is continuously unstable | The sensor window is attached by a foreign object | Cleaning the sensor window surface |
| Sensor self-cleaning damage | Replace the cleaning brush |
This product includes:
The company provides local after-sales service within one year from the date of sale, but does not include damage caused by improper use. If repair or adjustment is required, please return it, but the shipping cost must be conceited. Damaged on the way, the company will repair the damage of the instrument for free.