RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision

  • RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision
  • RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision
  • RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision
RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision
Product Details:
Place of Origin: CHINA
Brand Name: kacise
Certification: CE
Model Number: KFDO601
Payment & Shipping Terms:
Minimum Order Quantity: 0-100
Price: $0-$2000
Packaging Details: Common package or custom package
Delivery Time: 3-10days
Payment Terms: L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability: 100
Contact Now
Detailed Product Description
Measuring Principle: Fluorescence Method Range Ability: 0~20mg/L (0 ≤ 200% Saturation)
Resolution Ratio: 0.01mg/L,0.1℃ Precision: ±2%F.S.,±0.5℃
Temperature Compensation: Automatic Temperature Compensation(Pt1000) Output Mode: RS-485 Bus, Modbus-RTU Protocol
High Light:

RS485 Interface Dissolved Oxygen Sensor

,

Anti Interference Dissolved Oxygen Sensor

,

316L Stainless Steel Dissolved Oxygen Probe

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601

KFDO601 Online Fluorescence Dissolved Oxygen No Electrolyte Is Required And Will Not Be Polarized No Oxygen Consumption

 

 

1.Difference between polarographic and fluorescence measurements of dissolved oxygen

 

 

1: Fluorescence method dissolved oxygen measuring instrument is based on the principle of fluorescence quenching. The fluorescent material at the front of the sensor is a special platinum porphyrin composite of a polyester foil that allows the gas molecules to pass through. The surface is coated with a black light-blocking material to prevent the interference of sunlight and other fluorescent substances in the water.

The blue light illuminates the fluorescent material and causes it to excite and emit red light. Since oxygen molecules can take away energy (quenching effect), the time and intensity of the excited red light is inversely proportional to the concentration of oxygen molecules. By measuring the phase difference between the excitation red light and the reference light, and comparing with the internal calibration value, the concentration of oxygen molecule can be calculated. After linearization and temperature compensation, the final value is output.

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 0

2: The polarographic sensor consists of a silver anode and a gold cathode with a ring at the bottom. A thin semi-permeable film, spread over the sensor, can isolate the electrode from the outside while allowing gas to enter. The bottom of the sensor is filled with a small amount of surfactant electrolyte to enhance wetting during operation.

When a polarization voltage is applied to the electrode of the polarographic sensor, oxygen penetrates the membrane and reacts at the cathode, generating a current.

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 1

 

 

2.Feature

 

  • No electrolyte is required and will not be polarized.
  • No oxygen consumption, not affected by flow rate.
  • Built-in temperature sensor, automatic temperature compensation.
  • Free from interference by chemicals such as sulfides
  • Small drift, rapid reaction, more accurate measurement.
  • Long service cycle, lower cost of use.
  • Simple replacement of fluorescent cap.
  • Rs-485 interface, Modbus/RTU protocol.
  • Low power consumption,anti-interference.

 

3.Technical Specifications

 

Model KFDO601
Measuring principle Fluorescence method
Range ability 0~20mg/L (0 ≤ 200% saturation)
Resolution ratio 0.01mg/L,0.1℃
Precision ±2%F.S.,±0.5℃
Temperature compensation Automatic temperature compensation(Pt1000)
Output mode RS-485 bus, Modbus-RTU protocol
Working conditions 0~45℃,<0.2MPa
Storage temperature -5~65℃
Installation mode Immersion mounting
Cable length 5 meters, other lengths customizable
Power dissipation <0.3W@12V
Source 12~24VDC ±10%
levels of protection IP68
Calibration Two-point calibration
Fluorescent cap life Guaranteed use for one year (under normal use)
Sensor shell material POM and 316L stainless steel
 

4.Dimensional Drawing

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 2

Note:The sensor connector is m16-5 core waterproof connector male.

 

5.Installation and Electrical Connection

 

5.1 Install

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 3

 

 

The sensor shall be installed fixedly below the liquid level. The installation and use shall avoid collision or scraping against the surface of the fluorescent film head. The fluorescent film head part shall avoid being attached by the bottom sediment.Remove the rubber cover when use.

5.2 Electrical connection

  • Red wire-power cord(12~24 V)
  • Black wire-ground wire(GND)
  • Blue Line-485 A
  • White Line-485 B

Check wiring sequence carefully before power on to avoid unnecessary losses caused by wiring errors.

Wiring instructions: Considering that the cables have been immersed in water (including seawater) or exposed to air for a long time, all wiring positions are required to be waterproof, and the cables should have certain corrosion resistance.

 

6.Maintenance

 

6.1 Maintenance Schedule and Methodology

6.1.1 Maintenance Schedule

Different from the dissolved oxygen probe technology of electrochemical principle, the fluorescence dissolved oxygen probe does not consume oxygen and does not need to be cleaned frequently (except when it is used in viscous liquid).

Maintenance Task Recommended Maintenance Frequency
Cleaning sensor Wash every 30 days
Check for damage to sensors and fluorescent caps Check every 30 days
Replace the fluorescent cap Replace it once a year

Calibrate sensors (if required by the competent authority))

According to the maintenance schedule required by the competent department
 

Note: The maintenance frequency in the above table is only recommended, and the maintenance personnel shall clean the sensor according to the actual use of the sensor; however, the replacement frequency of the fluorescent cap is recommended once a year.

6.1.2 Maintenance Method

a) Sensor outer surface: clean the outer surface of the sensor with tap water, if there is still debris residue, wipe with wet soft cloth, for some stubborn dirt, you can add some household washing liquid to tap water to clean.

b) Surface of fluorescent film head: If there is dirt on the surface of fluorescent film head, please rinse with clean water or wipe gently with a soft cloth.Pay attention to the intensity of cleaning to avoid causing scratches in the measuring area and affecting the measurement accuracy.

c) Inside the fluorescent film head: it is generally unnecessary to clean. If water vapor or dust enters into the fluorescent film head, the cleaning steps are as follows:

  • Screw down the fluorescent film head;
  • Rinse the inner surface of the fluorescent film head and the optical window of the sensor.
  • For oily dirt, you can use household cleaning solution;
  • Gently dry the water with a clean flannelless cloth and let dry;
  • Reinstall the fluorescent film head.

d) Check the cables of the sensor: there should be no damage to the skin and root of the cables;Terminals should not be submerged in water;When the sensor is normally installed, the cable should not be tensioned, otherwise the internal wire of the cable will easily break and the sensor cannot work normally.

e) Check whether the casing of the sensor is damaged by corrosion or other causes.

f) Daily storage of fluorescent film head: When not in use, a rubber protective cover with a wet sponge should be covered to keep the surface of the measuring area of fluorescent film head wet.If the surface of the measuring area of the sensor fluorescent film head is chronically dry, measurement error or data instability will occur, and it shall be soaked in water for 48 hours before use.

6.2 Frequently Questions

Wrong Probable Cause Solution
The operating interface cannot connect or does not display the measurement results Error connecting controller to cable Reconnect the controller and cable
Cable failure Please contact us.
The fluorescent cap is not tightened or damaged Refit and tighten the fluorescent cap or replace the fluorescent cap.
The measured value is too high, too low, or the numerical value remains unstable. The outer surface of the fluorescent cap is attached to the outer object Clean the outer surface of the fluorescent film head and agitate the film head during measurement.
The fluorescent cap was damaged Replace the fluorescent cap

The fluorescent cap has exceeded its service life

Temperature measurements change slowly. The temperature measuring area (stainless steel housing) is attached to the exterior Gently brush the attachment with a soft brush

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6.3 Calibration of Sensors

a) Zero Calibration

Weighing 5 g of sodium sulfite by a balance, adding 95mL of water into a 250-mL measuring cylinder, pouring the water into a beaker, adding the sodium sulfite which has been weighed, stirring with a glass rod, dissolving, and obtaining a solution of 5% sodium sulfite, putting the sensor in a solution, And the zero point calibration is carried out after the three-minute numerical stability is stable. Refer to the Appendix to the instructions

b) Slope Calibration

The sensor probe is placed in air saturated water and the slope is calibrated after 3 minutes of numerical stability. The instructions refer to the appendix.

c) Preparation of Air Saturated Water: add 2% 3 volume fresh distilled water to the constant temperature water bath to float the porous plastic sheet on the water surface (see figure below). At the same time, the bubbler (air pump) is used to aerate the water continuously for more than 1 hour, stop aeration, and get air saturated water after 20 minutes or so. Put the sensor into the water and calibrate the slope after the numerical value is stable.

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Note: as an option, slope calibration can also be performed in water-saturated air. Put the sensor in a calibration bottle with a small amount of water (the probe is higher than the water surface 2-3mm) to ensure that the sensor film cap remains wet but has no water droplets, and the calibration slope is calibrated after 3 minutes of numerical stability.

6.4 Points for Attention

  • Avoid sun exposure to the inner surface of the fluorescent cap.
  • Please don't touch the fluorescent film with your hands.
  • Measuring and calibrating the surface of fluorescent film to avoid attaching bubbles.
  • Avoid directly applying any mechanical stress (pressure, scratches, etc.) to the fluorescent film in use.

 

6.Company Information

 

 

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 5

Xi'an Kacise is certified by ISO 9001-2008, GJB 9001B-2009, CE, intrinsic safety explosion-proof and isolated safety explosion-proof, etc. which assure our products are conform to the international standards.

We have been specializing in manufacturing pressure, level and flow measuring instruments and related products for 10 years. We have experienced technical and management personnel, through the efforts of our entire staff, we have become a professional measuring instrument manufacturer and solution provider. And we are continuously trying to improve product quality.

Currently, Kacise sensor is already exported to America, Spain, Australia, Chile and 20 other countries and regions, and received many favorable feedbacks. Our overseas distributers are also increasing. We regards “reasonable prices, efficient production time and good after-sales service” as our tenet, and hope to cooperate with more customers for mutual development.

 

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RS485 Interface Dissolved Oxygen Sensor Anti Interference KFDO601 with precision 7

 

 

 

 

 

 

 

 

 

Contact Details
Xi'an Kacise Optronics Co.,Ltd.

Contact Person: Ms. Evelyn Wang

Tel: +86 17719566736

Fax: 86--17719566736

Send your inquiry directly to us (0 / 3000)

Contact

Address: i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.

Factory Address:i City, No11, TangYan South road, Yanta District, Xi'an,Shaanxi,China.