Product Details
Output:
RS-485(Modbus/RTU)
Power Consumption:
5W@12V
Power Supply:
12VDC±5%
Accuracy:
±0.1pH
Measuring Range:
0~100mg/L Or 0~1000mg/L
Highlight:

Multi-parameter Water Quality Sensor

,

8 Parameters Water Quality Sensor

Payment & Shipping Terms
Minimum Order Quantity
1PCS
Packaging Details
each unit has individual box and all boxes are packed in standard packages or customers requests available
Delivery Time
5-8 work days
Payment Terms
L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability
1000 Piece/Pieces per Week negotiable
Product Description
KWS-850 On-line Multi-parameter Water Quality Sensor

This advanced sensor simultaneously measures 8 critical water quality parameters, providing comprehensive monitoring for industrial, environmental, and municipal applications.

Key Specifications
Output RS-485(Modbus/RTU)
Power Consumption 5W@12V
Power Supply 12VDC±5%
Accuracy ±0.1pH
Measuring Range 0~100mg/L or 0~1000mg/L
Product Features
  • Integrated design measures 8 parameters simultaneously (including temperature)
  • Digital sensor with RS-485 bus and Modbus/RTU communication protocol
  • Automatic cleaning device prevents microorganism buildup for accurate measurements
  • Optional digital sensors for dissolved oxygen, COD, conductivity/salinity, turbidity, ammonia nitrogen, pH, and ORP
  • Ideal for long-term online monitoring applications
Technical Parameters
Dissolved Oxygen Probe
Measuring Range 0~20mg/L
Accuracy ±0.4mg/L
Resolution 0.01mg/L
Turbidity Probe
Measuring Range & Accuracy 0~100NTU | ±3% or ±2NTU
0~1000NTU | ±5% or ±3NTU
Resolution 0.1NTU
Conductivity/Salinity Probe
Measuring Range & Resolution 0~5000uS/cm | 1uS/cm
0~200mS/cm | 0.1mS/cm
0~70PSU | 0.1PSU
Accuracy ±1.5% F.S.
COD Probe
Measuring Range COD | Turbidity
0~200mg/L equiv. KHP | 0~100NTU
0~500mg/L equiv. KHP | 0~200NTU
COD Accuracy ±5% F.S.
COD Resolution 0.1mg/L
Turbidity Accuracy ±5% F.S.
Turbidity Resolution 0.1NTU
pH Probe
Measuring Range 0~14pH
Accuracy ±0.1pH
Resolution 0.01pH
ORP Probe
Measuring Range -1500mV~+1500mV
Accuracy ±6mV
Resolution 1mV
Ammonia Nitrogen Probe
Measuring Range 0~100mg/L or 0~1000mg/L
Accuracy ±10% or ±2mg/L
Resolution 0.1mg/L
Temperature
Measuring Range 0~50℃
Accuracy ±0.5℃
Resolution 0.1℃
Additional Specifications
Cleaning Method Automatic cleaning
Cable Length 5 meters (custom lengths available)
Wiring Information
Red Line Power Supply (VCC)
White Line 485 Data _B (485_B)
Blue Line 485 Data _A (485_A)
Black Line Ground Line (GND)
Installation Diagram
The KWS-850 On-line Multi-parameter Water Quality Sensor Can Measure 8 Parameters Simultaneously 0
Installation Note: Use lifting rings or pipe thread installation to avoid direct force on the cable. The sensor connector is M16-5 core waterproof male connector.

Send your inquiry directly to us

Privacy Policy China Good Quality Fluid Level Meter Supplier. Copyright © 2018-2025 Xi'an Kacise Optronics Co.,Ltd. . All Rights Reserved.