Brand Name:
Yokogawa
Model Number:
DO30G
| Parameter | DO30G |
|---|---|
| Object of Measurement | Concentration of oxygen dissolved in solution (water) |
| Principle of Measurement | Galvanic-cell method |
| Measurement Range | 0 to 20 ppm, 0 to 20 mg/L, or 0 to 100% (saturation) The measurement range must be entered through the dissolved oxygen converter. |
| Resistance Temperature Detector (RTD) | Pt1000 |
| Construction | Sensor unit: Rigid PVC, Stainless steel O-ring: Nitrile rubber Permeable membrane: Fluoro-resin [FEP] Cable shielding: Heat-resistant flexible PVC Outlet of cable: Polycarbonate resin |
| Sensor Cable | Length: 3 m, 5 m, 10 m, 15 m, or 20 m (specified when ordering) Crimping terminal configuration: Pin-type, M4/M3 Ring-type or Fork-type (specified when ordering) |
| Weight | Approx. 0.3 kg + 0.12 × N kg (N: cable length, e.g., for cable length of 5 m, approx. 0.9 kg) |
| Temperature of Solution Under Measurement (SUM) | max. 0 to 40 °C |
| Flow rate of SUM | No less than 20 cm/s Recommended to prevent sensor from trapping bubbles that cause measurement errors |
| Pressure of SUM | Normally: 0 to 30 kPa Maximum applicable: 100 kPa |
| Characteristics | Repeatability: 0.1 mg/L or 3% FS, whichever is greater (including sensor) Temperature Compensation Error: Within ±3% FS (including sensor for a ±5 °C change in the range of 0 to 40 °C) Response Time: Within 2 minutes (90% response) (including sensor) |
Send your inquiry directly to us