Model Number:
KOP601
Used in environmental water quality monitoring, acid/alkali/salt solution, chemical reaction process, industrial production process, it can meet the requirements of online ORP measurement for most industrial applications.
| Model | KOP601 |
| Measuring Range | -1500~+1500mV |
| Resolution | 1mV |
| Accuracy | ±6mV |
| Working temperature | 0~65℃ |
| Working pressure | <0.1MPa |
| Power supply | 12~24VDC ±10% |
| Signal output | RS-485(Modbus/RTU) |
| Wetted material | POM |
| Installation | Immersion mounting, 3/4 NPT thread |
| Cable length | 5 meters, other lengths can be customized |
| Calibration | One point calibration |
| Power consumption | <0.3W@12V |
| Protection grade | IP68 |
![]()
Note:The sensor connector is m16-5 core waterproof connector male.
![]()
Note: The sensor should not be installed upside down or horizontally when installed, at least at an angle of 15 degrees or more.
After wiring is completed, it should be carefully checked to avoid incorrect connections before powering up.
Cable specification: Considering that the cable is immersed in water (including sea water) for a long time or exposed to the air, the cable has certain corrosion resistance. The outer diameter of the cable is Φ6 mm and all interfaces are waterproof.
When measuring the ORP electrode, it should be cleaned in distilled water (or deionized water) first, and the filter paper should be used to absorb moisture to prevent impurities from being introduced into the liquid to be tested. 1/3 of the electrode should be inserted into the solution to be tested.
The electrode should be washed when not in use, inserted into a protective cover with a 3.5 mol/L potassium chloride solution, or inserted into a container with a 3.5 mol/L potassium chloride solution.
Check if the terminal is dry. If it is stained, wipe it with absolute alcohol and dry it. Avoid long-term immersion in distilled water or protein solution and prevent contact with silicone grease. With a longer electrode, its glass film may become translucent or with deposits, which can be washed with dilute hydrochloric acid and rinsed with water. The electrode is used for a long time. When a measurement error occurs, it must be calibrated with the meter to correct it.
When the calibration and measurement cannot be performed while maintaining and maintaining the electrode in the above manner, the electrode has failed. Please replace the electrode.
2.1 g of hydrazine was weighed into 200 mL of a standard pH buffer solution having a pH of 4.00 and mixed.
Weigh 2.1 g of hydrazine to 200 mL of a standard pH buffer solution with a pH of 6.86 and mix.
Weigh 2.1 g of hydrazine to 200 mL of a standard pH buffer solution with a pH of 9.18 and mix.
Note that when preparing the pH, a total of 250 mL is prepared, and 200 mL is used for use.
pH to prepare the corresponding standard solution method:
Take 250mL of distilled water from a measuring cylinder, pour it into a beaker, add a packet of calibration powder with pH=6.86/4.00/9.18, stir evenly with a glass rod until the powder is completely dissolved, and prepare a solution with pH=6.86/4.00/9.18.
Place the sensor in the prepared 86mV (or 256mV or -40mV) solution and wait for 3 to 5 minutes. After the value is stable, see if the displayed value is 86mV (or 256mV or -40mV). If not, calibration is required. See the appendix for calibration instructions.
This product includes:
The company provides local after-sales service within one year from the date of sale, but does not include damage caused by improper use. If repair or adjustment is required, please return it, but the shipping cost must be conceited. Damaged on the way, the company will repair the damage of the instrument for free.