KEC310 Online Conductivity Sensor Easy to install submerged installation Ip68

KEC310 Online Conductivity Sensor Easy to install submerged installation Ip68

Place of Origin:

CHINA

Brand Name:

kacise

Certification:

CE

Model Number:

KEC310

Contact Us
Request A Quote
Product Details
Range And Resolution:
0-5000μs/cm 1
Precision:
± 1.5% F.S.
Operating Temperature:
0 ~ 65 °C
Working Pressure:
< 0.6 Mpa
Power Supply:
12 ~ 24 VDC ± 10%
Signal Output:
RS-485(Modbus/RTU)
Payment & Shipping Terms
Minimum Order Quantity
0-100
Price
$0-$2000
Packaging Details
Common package or custom package
Delivery Time
3-10days
Payment Terms
L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability
100
Product Description

1.Application Environment

Drinking water/surface water/various water supplies/industrial water treatment

2.Feature

  • Signal Output: RS-485(Modbus/RTU protocol)
  • Convenient to connect to PLC, DCS, Industrial Control Computer, General Controller, paperless recording instrument or touch screen and other third-party equipment.
  • Easy to install: submerged installation.
  • Ip68, containment level.

3.Technical Specifications

Model number KEC310
Range and resolution 0-5000μs/cm 1
Precision ± 1.5% F.S.
Operating temperature 0 ~ 65 °C
Working pressure < 0.6 mpa
Power supply 12 ~ 24 VDC ± 10%
Signal output RS-485(Modbus/RTU)
Contact material Abs
Installation mode Immersion mounting
Cable length 5 meters, other length can be customized
Temperature compensation Automatic temperature compensation (PT1000)
Calibration mode Two-point calibration
Power consumption < 0.05 W
Protection level IP68
 

 

 

4.Dimensional Drawing

KEC310 Online Conductivity Sensor Easy to install submerged installation Ip68 0

5.Installation and Electrical Connection

5.1 Installs

KEC310 Online Conductivity Sensor Easy to install submerged installation Ip68 1

Note: At least 2 cm away from the bottom and side walls of the container during installation and testing.

5.2 Electrical connection

The cable is a 4-core twisted pair shielded wire defined in line sequence:

Red wire -- power cord (12V-24VDC)

Black line (GND)

Blue Line -- 485a

White line -- 485b

Before electrification, the wiring sequence should be carefully checked to avoid unnecessary losses caused by wiring errors.

Wiring instructions: Consider the cable long-term immersion in water (including seawater) or exposed to air, all wiring is required to do waterproof treatment, customer cable should have a certain anti-corrosion ability.

 

6.Maintenance

6.1 Use and Maintenance

The regular electrode needs to be cleaned and calibrated periodically, and the maintenance cycle is decided by the customer according to his own working condition. Conventional electrode cleaning method: with a soft brush to remove the attachment (care to avoid scratching the electrode surface) , and then clean with distilled water, after the calibration operation.

Cleaning method of inductive electrode:

Inductive electrode is basically maintenance-free; the shell of pollution or mild scaling does not affect its normal work.

If cleaning, with a soft brush or sandpaper to remove the attachment, and then use distilled water cleaning, after the calibration operation.

Because inductance type electrode often works in easy to scale up or dirty environment, cleaning can be appropriate to increase the force. A minor scratch on the electrode surface does not affect the normal operation of the electrode, but to avoid penetrating the electrode shell.

6.2 Calibration

a) Zero Calibration

Rinse the sensor with distilled water and drain the liquid with filter paper. The sensor is connected to the power supply and placed vertically in the air for about 3 minutes. After the numerical value is stable, zero calibration is carried out. The calibration instructions are detailed in the appendix.

b) slope calibration

Place the electrode vertically in the standard solution (20% full range) and note that the slope of the electrode is calibrated at least 2 cm from the bottom and side walls of the container. The calibration instructions are detailed in the appendix.

7.Quality and Service

7.1 Quality Assurance

Quality inspection departments have standardized inspection procedures, with advanced and perfect testing equipment and means, and in strict accordance with the rules of inspection, 72-hour aging of products to do experiments, stability experiments, do not let a unqualified product factory.

The consignee will return the lot of the product with 2% nonconformity rate, and all the cost will be borne by the supplier. The standard of consideration refers to the product description provided by the supplier.

Guarantee the quantity and speed of delivery.

7.2 Spare Parts

This product includes:

1 sensor

1 copy of instruction manual

Certificate 1

Calibration fluid (optional)

7.3 After-sales Service Commitment

Our company provides after-sale service for one year from the date of sale, but does not include damage caused by improper use. If repair or adjustment is needed, please return it at your own expense, the company will repair the damage of the instrument free of charge.

 

 