logo
Home > Products > Gas Detector Sensor >
100ppm Range SF6 Sensor High Efficiency Sulfur Hexafluoride Sensor with 10.6um Infrared Absorption

100ppm Range SF6 Sensor High Efficiency Sulfur Hexafluoride Sensor with 10.6um Infrared Absorption

100ppm Sulfur Hexafluoride Sensor

100ppm SF6 Sensor Module

Sulfur Hexafluoride SF6 Sensor Module

Place of Origin:

shanxi xi'an

Brand Name:

Kacise

Model Number:

KSS525

Contact Us
Request A Quote
Product Details
Range (customizable):
100ppm
Measurement Accuracy:
± 0.5ppm Or ±2% True Value Is The Greater Ppm
Storage Temperature:
-20-80℃
T90 Of Pump-suction:
5S
Operating Temperature:
-20-60℃
Repeatability Of 50%F.S.:
<+2%.
Work Pressure:
0.8-1.2 Atm
Life Span:
10 Years
Operating Humidity:
0-90%RH
Supply Voltage:
12V
Maximum Operating Current:
120-150mA
Highlight:

100ppm Sulfur Hexafluoride Sensor

,

100ppm SF6 Sensor Module

,

Sulfur Hexafluoride SF6 Sensor Module

Payment & Shipping Terms
Minimum Order Quantity
1
Packaging Details
each unit has indicidual box and all boxes are packed in standard packages or customers requests available
Delivery Time
5 work days
Payment Terms
T/T,Western Union,MoneyGram
Supply Ability
1000 Pieces per Week
Product Description

KSS525 Infrared SF6 Gas Sensor Module

1. Overview

KSS525 is a gas sensor module using NDIR infrared absorption detection principle, which determines the concentration according to the absorption of SF6 gas at 10.6um. KSS525 adopts MEMS light source, optical cavity with special structure and dual-channel detector to realize spatial dual-optical path reference compensation, and microprocessor for signal acquisition, processing and output. KSS525 has good selectivity, high sensitivity, no oxygen dependence, long life, low power consumption; Built-in temperature sensor for temperature compensation; At the same time, it has UART, 485 output and analog voltage output, which is convenient for customers to various applications.

KSS525 physical image

100ppm Range SF6 Sensor High Efficiency Sulfur Hexafluoride Sensor with 10.6um Infrared Absorption 0
100ppm Range SF6 Sensor High Efficiency Sulfur Hexafluoride Sensor with 10.6um Infrared Absorption 1
2. Technical parameters
parametersymbolminimumTypical valuesmaximumunit
Storage TemperatureTstg-20-80°C
Operating TemperatureTA-2060°C
Operating HumidityHA090% RH
Work PressurePA0.81.2atm
Supply VoltageVs91224V
Maximum operating currentIavg100120150mA
Range (customizable)100ppm
Measurement accuracyAcurracy0.5ppm or ±2% true value is the greater%vol
T90Pump suction5second
Repeatability50%F.S.<+2%.
Life Span10year
3. Structural dimension drawing
100ppm Range SF6 Sensor High Efficiency Sulfur Hexafluoride Sensor with 10.6um Infrared Absorption 0
100ppm Range SF6 Sensor High Efficiency Sulfur Hexafluoride Sensor with 10.6um Infrared Absorption 1
4. Installation instructions

The sensor is mounted with a positioning hole spacing of 64 mm and an aperture of 3 mm
The wiring socket spacing is 2.54 m m

5. Matters that should be paid attention to in maintenance
  • The sensor should be calibrated regularly, it is recommended that it be no more than 3 months, and calibration is not required if automatic calibration is turned on for long-term operation
  • Do not use the sensor for a long time in an environment with a high density of dust
  • Please use the sensor within the range of the sensor's power supply
6. Ordering instructions

In order to be able to purchase the specific specifications required, please provide the following information

  1. Sensor name
  2. Measure the gas
  3. Sensor range
  4. For example, SW521R200, SF6, 0-100 ppm
7. Release Notes

In order to facilitate the better use of this product, V1.2 updates the calculation method of returning concentration value in the 4.3.2 MODBUS RTU protocol and the command description of reading concentration value after changing the address.

8. Ordering Information
Order information sheet
KSS525KSS525 infrared SF6 gas sensor module
xxxxThe sensor measures the range of CO2 concentration in ppm, with a minimum value of 2000 and a maximum value of 50000 ppm.
2000Range 200ppm (default).
10000Range 10000ppm
50000Range 50000ppm
encodeThe reaction speed is divided into two types: fast and slow
SSlow (default).
Qfast
encodeBaud rate selection, support commonly used baud rate 2400 9600 19200 38400bps, 8 bits of data, 1 bit of stop bit, no check bit: Confirm special needs before ordering.
CustomConfirm the baud rate before ordering
24002400bps baud rate
96009600bps baud rate
1920019200bps baud rate
3840038400bps baud rate (default)
encodeSerial port protocol
Modbus-RTUStandard Modbus-RTU protocol (default).
Modbus-ASCIIStandard Modbus-ASCII protocol
Modbus-SelfModbus private protocol
KSS525-2000-S-38400-Modbus-RTU
Contact Details

URL: www.kacise.com
Tel: +86-29-17719566736
Email: sales@kacise.com
Address: Tangyan South Road, Xi'an City, Shaanxi Province, China

Appendix
ShorthandFull name
VOL1% VOL refers to 1% of the volume of a particular gas in air.
PPM1PPM means that the volume of a specific gas in the air accounts for one millionth.
O2Oxygen molecules
LCDLCD display
RS485Asynchronous serial port 485
DCdirect current
ACCommunication
PVCpolyvinyl chloride

Send your inquiry directly to us