Place of Origin:
CHINA
Brand Name:
KACISE
Certification:
CE
Model Number:
KCF6021-1-900mm
The KCF600 series capacitive level transmitter is a perfect solution for continuous detection of various liquids in harsh conditions such as high temperature and pressure, corrosion, crystallization, and blockage. This transmitter is ideal for measuring water, acid and alkali solutions, and the water level in a boiler. Unlike traditional level transmitters, this machine is free of any movable or flexible parts and has the unique characteristic of impact resistance, easy installation, high reliability, high precision, and reasonable cost.
The device incorporates advanced RF capacitive sensing circuit for precise compensation of temperature and linearity correction, which is then transformed into a standard electrical signal that usually ranges from 4 to 20 mA. The system configuration provides the option of HART, CANBUS, and 485 communication protocols. The transmitter's full range has a self-calibration feature, enabling users to automatically calibrate the "zero" and "range" values with the simple push of two keys. This is a highly adaptable device that is suitable for replacing traditional float type, input type, and differential pressure type of level transmitters in various situations.
High Reliability and Minimal Maintenance
Our products are designed with simple structures that do not have any movable or elastic components. As a result, our products have high reliability and require minimal maintenance. Under normal circumstances, our products do not require regular medium and small repairs.
Signal Output Variety
We understand that different systems require different configurations. That's why our products come with a variety of signal output options to facilitate these different needs.
Accurate Measurements
Our products are equipped to handle high temperature and high pressure containers, and our measurements are not impacted by the temperature, specific gravity, or shape of the container, nor the influence of pressure. This makes our products highly reliable for precise measurements in extreme conditions.
Suitable for Corrosive Liquids
We understand that not all liquids are created equal. Our products are especially suitable for measuring corrosive liquids such as acids and alkaline solutions. You can trust our products to provide accurate and reliable measurements in these harsh conditions.
Overcurrent, Overvoltage, and Power Polarity Protection
We value the safety of our users. That's why our products come with the perfect protection against overcurrent, overvoltage, and power polarity issues. You can trust our products to provide safe and reliable operations in any situation.
The cable and rod level transmitter has a variety of fixed ways to adapt to different site conditions. When the site cannot provide the process facilities such as the M20 × 1.5 female thread or flange, you can choose three methods of installation (wall-mounted, horizontal pipe rack and vertical tube rack, which make you handy in the on-site installation.(With pictures)
![]()
![]()
When measuring the media of the larger mobility, the user can choose the weight attachment in order to prevent the sensor form violent swing, which can play the role of the stability of the sensor. The heavy objects must be put into the bottom of the container when adding weight, so as not to break the sensor’s wire.
When the site conditions are bad, the user should use the sensor shield to protect the sensor in order to avoid foreign body damage or affect the measurement accuracy.
If the containers are not metal, we need to put the transmitter enclosure with a wire access to the bottom of the container when using a KCF600 rod level transmitter; otherwise the measurement is not accurate.
This product has been calibrated at the factory; he need for field adjustment may take the following methods:
A: Intelligent circuit debugging method:
Site-zero----- When the liquid level is located in the lowest place, holding down the "Z" key and "S" key for 8 seconds after the same time release, and then "Z" key click, the output current is automatically adjusted to 4mA.
Site - filled ------ When the liquid level at the highest place, holding down the "Z" key and "S" key for 8 seconds after the same time release, and then "S" key click ,the output current is automatically adjusted to 20mA.
B: Analog circuit debugging method:
When the liquid level in the next limit, use the screwdriver slowly rotating "Z" potentiometer, observe the ammeter readings close to 4mA
When the liquid level in the upper limit, use the screwdriver slowly rotating "S" potentiometer and observe the ammeter readings close to 20mA
And operate it for 2-3 times more, until the zero and span were accurate 4mA and 20mA. After adjustment, tighten the cover of the transmitter should be.
The Kacise Fluid Level Meter is made of high-quality 316 Stainless Steel or PTFE material that ensures durability and long-lasting performance. It is suitable for use in various scenarios and applications such as Capacitive Oil Level Sensor for Truck and Industrial Tanks.
This product has an accuracy of ±2%, ±5% and a temperature drift of ≤0.02%FS/℃ (Range of 0~70℃). It can withstand storage temperatures of -55℃~+125℃ and probe temperatures of -50~250℃ making it ideal for use in harsh industrial environments.
The Kacise Fluid Level Meter is perfect for a wide range of applications such as monitoring the fuel level in trucks, measuring the oil level in industrial tanks, and other fluid level measurement applications. Its capacitive fuel sensor technology ensures accurate and reliable measurements in all scenarios.
Our Fluid Level Meter has a power supply voltage of 2~28VDC (and should be powered by the safety barrier). The probe temperature range is -50~250 ℃ with an accuracy of ±2%, ±5%. The range for this product is 0.01~30 M and it has a storage temperature range of -55℃~+125 ℃.
Need a custom Capacitive Oil Level Sensor for your truck or industrial tank? Look no further than Kacise! Contact us today to learn more about our product customization services.
Product Packaging:
The Fluid Level Meter product will be packaged securely in a box with proper cushioning to prevent any damage during transit. The package will also include an instruction manual and any necessary installation hardware.
Shipping:
The Fluid Level Meter product will be shipped via standard ground shipping. Customers can expect the product to be delivered within 5-7 business days after placing the order.
Send your inquiry directly to us